Technical Documentation

Technical documentation is the evidence package that shows a device meets the IVDR. Notified bodies review it in depth for Class B, C and D devices, and authorities can request it at any time.

What Annex II requires

  • Device description and specification: intended purpose, users, principles of the method, components, variants, previous generations and similar devices
  • Information supplied by the manufacturer: labels and instructions for use in the languages of the target markets
  • Design and manufacturing information: design stages, manufacturing processes and sites, including key suppliers
  • General safety and performance requirements: a checklist showing how each Annex I requirement is met, the standards and common specifications applied, and where the evidence sits
  • Benefit-risk analysis and risk management
  • Product verification and validation: analytical and clinical performance, stability, software verification and validation, and performance evaluation reports

Annex III adds the post-market surveillance documentation: the PMS plan, PMPF plan and the resulting reports.

Tips for a strong file

  • Structure the file exactly as Annex II is structured so reviewers can navigate it
  • Use the GSPR checklist as the map to every piece of evidence
  • Keep a single, consistent intended purpose statement across labeling, risk, performance and marketing material
  • Include clear rationales, not just test reports
  • Control versions carefully; mismatched document versions are a frequent finding
  • Plan translations of labeling early

IVDR guide
Overview  ·  Scope and Key Definitions  ·  Risk Classification  ·  Conformity Assessment and Notified Bodies  ·  Performance Evaluation  ·  Technical Documentation  ·  Post-Market Surveillance and Vigilance  ·  Economic Operators, UDI and EUDAMED  ·  Transition Deadlines and Reform

Need help? Salto Partners guides diagnostics, genomics and medical device companies through ISO 13485 and IVDR. Send us an email and we will set up a conversation.

For general information only; not legal or regulatory advice. Regulations and deadlines change, so confirm current requirements with your notified body or regulatory counsel. Last updated October 2026.

© 2026 Salto Partners Inc. · Strategy. Execution. Compliance.